Hanna Instruments Digital Refractometer for Ethylene Glycol Analysis
The HI96831 is a rugged, portable digital refractometer designed for the measurement of percent volume and freezing point of ethylene glycol based solutions. The HI96831 has a high accuracy of ±0.2% volume and ±0.5° C freezing point and features simple operation that gives reliable results each and every time. All readings are automatically compensated for temperature variations and displayed within a 1.5 second response time. The sealed flint glass prism and stainless steel well are easy to clean. Just wipe with a soft cloth in preparation for the next sample.
Sample size as small as 2 metric drops (100 μl)
Sealed stainless steel well with high-grade optical prism made of flint glass
Fast 1.5 second response time for temperature compensated readings
The HI96831 Digital Refractometer utilizes the measurement of refractive index to determine the percent volume and freezing points of ethylene glycol based solutions. Ethylene glycol is used in glycol cooling systems for industrial applications. Maintaining the appropriate percent of glycol is important in glycol chillers as too low of a percentage could lead to the system freezing and rupturing while too much glycol will reduce the systems efficiency. The instrument converts the refractive index of a sample to percent volume or freezing point. These conversions are based on internationally recognized references for unit conversion and temperature compensation for glycol solutions (e.g. CRC Handbook of Chemistry and Physics, 87th Edition). The HI96831 is an easy-to-use tool for measuring freezing point and percent volume in ethylene glycol solutions, in the field or in the lab.
Designed for ethylene glycol analysis
Automatic Temperature Compensation
0 - 100% Volume range with ±0.2% accuracy
0 to -50 °C freezing point range with ±0.5 °C accuracy
Making a Standard Solution
To make an Ethylene Glycol Solution, follow the procedure below:
Place container (such as a glass vial or dropper bottle that has a cover) on an analytical balance.
Tare the balance.
Volumetrically add X mL of high purity ethylene glycol (CAS #: 107-21-1; MW 62.068) to a 100 mL Grade A volumetric flask.
Use distilled or deionized water to bring the flask close to the total volume, mix, allow the solution to return to room temperature.
Once the solution has returned to room temperature use distilled or deionized water to bring the total volume to 100 mL. Mix solution well before use.
Ethylene Glycol
Total Volume
Expected Freezing Point Value
10 % V/V
10.00 mL
100.00 mL
-3.8 °C (25.2 °F)
40 % V/V
40.00 mL
100.00 mL
-26.3 °C (-15.4 °F)